• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CELESTICA ELECTRONICS S PTE LTD BIS; ELECTRODE, CUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CELESTICA ELECTRONICS S PTE LTD BIS; ELECTRODE, CUTANEOUS Back to Search Results
Model Number 186-0106
Device Problem Low Readings (2460)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/30/2022
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a procedure, the patient was awake and the sensor was showing a reading of 40. there was no patient injury.
 
Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the bis monitor reported lower than expected readings.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIS
Type of Device
ELECTRODE, CUTANEOUS
Manufacturer (Section D)
CELESTICA ELECTRONICS S PTE LTD
no. 6 serangoon north avenue 5
singapore 55491 0
SN  554910
Manufacturer (Section G)
CELESTICA ELECTRONICS S PTE LTD
no. 6 serangoon north avenue 5
singapore 55491 0
SN   554910
Manufacturer Contact
tanja beshear
5920 longbow drive
boulder, CO 80301
3035306434
MDR Report Key14508336
MDR Text Key292969076
Report Number2936999-2022-00470
Device Sequence Number1
Product Code GXY
UDI-Device Identifier20884521134307
UDI-Public20884521134307
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K093183
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/05/2023
Device Model Number186-0106
Device Catalogue Number186-0106
Device Lot Number0105221K
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/10/2022
Initial Date FDA Received05/26/2022
Supplement Dates Manufacturer Received06/30/2022
Supplement Dates FDA Received07/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age47 YR
Patient SexMale
-
-