• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V3
Device Problem Material Perforation (2205)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/04/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the scope was leaking from the bending section due to a puncture found in the distal end.There was also a broken pin at the bending section skeleton that had punctured through the distal end, contributing to the aforementioned leakage.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
The customer originally returned her olympus uretero-reno videoscope due to the unit not displaying an image during an unspecified procedure.Additional details relating to the patient and the event have been requested, but no response has been received at this time.There was no patient harm or consequence reported as a result of this event.During the device evaluation, it was discovered that the bending section of the device was broken and leaking.This report is being submitted to capture the broken bending section found during the device evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The investigation confirmed the reported device problem.A review of the device history record found the subject device was refurbished, however the data regarding the refurbishment details is unavailable.A definitive root cause for the reported product problem was not established.However, it is probable that the issue occurred because the end user applied excessive force on the bending section; causing it to break.Design measures had previously been taken to address potential injury due to a kinked/cracked bending tube protruding during the procedure.The design of the bending tube was changed to not cause damage to the patient even if broken.As there was no reported patient injury due to the subject event, these measures are still considered effective.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14521977
MDR Text Key300593954
Report Number8010047-2022-09020
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403385
UDI-Public04953170403385
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/05/2022
Initial Date FDA Received05/27/2022
Supplement Dates Manufacturer Received06/24/2022
Supplement Dates FDA Received07/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-