• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ENSITE VELOCITY¿ STOCKERT GENCONNECT; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ENSITE VELOCITY¿ STOCKERT GENCONNECT; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 100036696
Device Problem High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/19/2022
Event Type  malfunction  
Event Description
Related manufacturer ref#2184149-2022-00134.During an atrioventricular reciprocating tachycardia procedure, an impedance issue occurred causing a delay.When using ensite precision and the atakr genconnect, the connection was unsuccessful.The rf generator was replaced with a stockert and the impedance baseline was high in the left atrium.The impedance baseline was checked without the genconnect and the values were 100 ohms less.The genconnect was replaced and the issue was resolved.The procedure was completed without patient consequences.
 
Manufacturer Narrative
One ensite velocity¿ stockert genconnect was received into the lab for investigation.Visual inspection of the returned product confirmed all input and output connectors are free of physical damage and all labeling is legible and correctly oriented.The returned product is manufactured specifically for use with the biosense webster radio frequency generator as the front panel label indicates.Functional testing performed verified circuit integrity on all channels during the evaluation period.Normal output amplitudes and impedance values were observed during functional testing.Based on the information provided to abbott and the investigation performed, the root cause of the reported event cannot be conclusively determined as no abnormalities were identified.The returned product operated as designed during the evaluation period.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY¿ STOCKERT GENCONNECT
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14545447
MDR Text Key300730006
Report Number2184149-2022-00133
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number100036696
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/19/2022
Initial Date FDA Received05/31/2022
Supplement Dates Manufacturer Received07/19/2022
Supplement Dates FDA Received08/08/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ENSITE VELOCITY¿ ATAKR GENCONNECT KIT
-
-