Analysis of the returned ipg revealed that the device could not be detected by a remote control after several charging attempts.An internal inspection found a completely depleted internal battery and excessive quiescent current leak.High voltage (vh) node to ground impedance also measured low suggesting that the device was damaged by high-voltage transients or high radio frequency(rf) energy, resulting in electrical shorts within the application-specific integrated circuit (asic).The electrical short in the asic caused the inability to communicate with the ipg battery.The cause of the asic damage is unknown.A product labeling review identified that the device was used per the instructions for use (ifu) product label.The instructions for use (ifu) cautions that the following medical therapies or procedures may turn the stimulation off, cause permanent damage to the stimulator or may cause injury to the patient: electrocautery, external defibrillation, diagnostic ultrasonic scanning, lithotripsy, radiation therapy, transcranial stimulation, x-ray and computed tomography scans.
|