• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. BIPOLAR HIP; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE46

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. BIPOLAR HIP; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE46 Back to Search Results
Model Number 412-02-046
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Unspecified Infection (1930)
Event Date 05/24/2022
Event Type  Injury  
Event Description
Revision surgery - the patient had an infection.Surgeon removed the ball and shell, then evaluated the stem and then did a thorough washout of the joint.Surgeon replaced the ball and shell with a new ball and shell.
 
Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 2 months and 3 weeks apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr # (b)(4) associated with the main part #412-02-046, shell, bipolar, modular, encore, size46 which document that out of 10 parts lot all the parts were rejected due to lot not found in oracle it can be a duplicated.Later, the rejected parts were reworked and accepted after proper justification.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR HIP
Type of Device
SHELL, BIPOLAR, MODULAR, ENCORE, SIZE46
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key14569820
MDR Text Key293180414
Report Number1644408-2022-00732
Device Sequence Number1
Product Code KWY
UDI-Device Identifier00888912077378
UDI-Public(01)00888912077378
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953510
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number412-02-046
Device Catalogue Number412-02-046
Device Lot Number991B1192A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2022
Initial Date FDA Received06/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
425-96-011 LOT 190R2147; 497-28-035 LOT 631C1406
Patient Outcome(s) Other; Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-