MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V3
Device Problem Degraded (1153)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Customer returned the device for the issue of leak observed during reprocessing.There is no patient involvement and no harm to any patient.Upon evaluation of the returned device, it was observed that the distal end bending section rubber coating (a-rubber) is porous and bending tube worn out.The bending section skeleton was defective.This medwatch is being submitted for the issues of a-rubber being porous, bending tube worn out, and defective bending section skeleton as observed during device evaluation.
 
Manufacturer Narrative
Investigation activities have been opened to manage the actions related to this issue and any required mdr reporting.The device is returned and an evaluation completed for it.Upon inspection of the device, it was observed that the distal end bending section rubber coating (a-rubber) is porous and bending tube worn out.The bending section skeleton was defective.In addition, it was noted that the a-rubber was leaking.The user reported issue of leak was confirmed.The connecting tube and cable tube unit had a scratch.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
 
Manufacturer Narrative
This report is being supplemented to provide additional information.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the investigation, possible causes may include the following: the user applied excessive stress to the bending section and the bending tube was broken.A definitive root cause cannot be identified.This issue is covered in the instructions for use: ·precautions : perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.Inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.Do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.Do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14594561
MDR Text Key300973321
Report Number8010047-2022-09379
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403385
UDI-Public04953170403385
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/21/2022
Initial Date FDA Received06/03/2022
Supplement Dates Manufacturer Received07/22/2022
Supplement Dates FDA Received08/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/14/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-