Model Number 279732440
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Device report from synthes reports an event in colombia as follows: it was reported that on an unknown date, the screwdriver showed rust on the inside and outside of the piece.The issue was identified before the equipment was sterilized.This report involves one expedium spine system quick connect reduction screwdriver.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Reporter is a j&j employee.Product code #: 279732440.Synthes lot #: mi21886.Supplier = micro pulse.Lot qty of 36 units were released on (b)(6) 2011 with no ncr generated during production.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the qc redux screwdriver had evidence of corrosion.The distal end of the device was also observed to be stripped.Light scratches and marks due to normal wear were observed on the device.No other damage or defects were noted.A dimensional inspection was unable to be performed due to the damage to the tip of the device, and it is irrelevant to the corrosion condition.The observed condition of the distal end of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the qc redux screwdriver would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM SPINE SYSTEM QUICK CONNECT REDUCTION SCREWDRIVER
Type of Device
SCREWDRIVER
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
SENTIO (INNOVATIVE SURGICAL SOLUTIONS, LLC)
50461 west pontiac trail
wixom MI 48393
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14594668
MDR Text Key301592274
Report Number1526439-2022-00775
Device Sequence Number1
Product Code HXX
UDI-Device Identifier10705034200248
UDI-Public(01)10705034200248
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number279732440
Device Catalogue Number279732440
Device Lot NumberMI21886
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/05/2022
Initial Date FDA Received06/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/27/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-