• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938739
Device Problems Fluid/Blood Leak (1250); Structural Problem (2506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2022
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the fill port of a 1000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) was wrongly assembled as it was not sealed inside the eva bag and as a result all the ingredients outflowed into the working environment.This was identified during infusing the parenteral nutrients into the bag.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured from january 16, 2020 - january 21, 2020.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph which observed that the fill port tube was attached to the outside, bottom, back of the bag.The reported condition was verified.The cause of the condition was due to incorrect assembly during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt 001 mz 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14597951
MDR Text Key293334980
Report Number1416980-2022-02897
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477237
UDI-Public(01)00085412477237
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/31/2022
Device Catalogue NumberH938739
Device Lot Number60222967
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/09/2022
Initial Date FDA Received06/03/2022
Supplement Dates Manufacturer Received07/06/2022
Supplement Dates FDA Received07/06/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-