• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL PRE-FILLED SALINE SYRINGE; SALINE VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL PRE-FILLED SALINE SYRINGE; SALINE VASCULAR ACCESS FLUSH Back to Search Results
Model Number 306546
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/06/2022
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: exec summary: no physical device samples were returned.The investigation was performed based primarily on the image(s) provided.It was reported the plunger is sticking.To aid in the investigation, two disassembled empty samples with no packaging blister or tip cap were received for evaluation by our quality team.A visual inspection was performed and no defects or imperfections were observed.Each sample was then tested for sustaining force, which is the force applied to the plunger rod while moving downwards when expelling the saline solution, and all results were within specification.It could be possible the customer is getting some products that are towards the high specification limit and are related to the symptom reported by the customer since they require extra force than normal to expel the solution.Previous investigations have revealed that improper silicone application within the syringe barrel can create plunger resistance.Several quality initiatives have been implemented on our manufacturing line to ensure that the silicone application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the silicone is uniformly applied to the syringe barrel.Based on the investigation and with the sample analysis the symptom reported by the customer could not be confirmed and the root cause could not be determined.A review of the complaint lot history check was performed and this is the 4th related complaint for plunger movement difficult on lot # 2021612.No non-conformance's were raised in association with this type of event for this lot, concluding all inspections were performed as per the applicable operations and met qc specifications.Capa/sa: based on the above no additional investigation and no corrective or preventative action (capa) or situational analysis (sa) are required at this time.Dhr review: a lot history review for batch 2021612 was carried out and no related non conformance's were raised in association with this packaged lot concluding all inspections were performed as per the applicable operations and met qc specifications.
 
Event Description
It was reported that 1 bd posiflush¿ normal pre-filled saline syringe had plunger movement issues.The following information was provided by the initial reporter : the customer reported that the plungers are stopping at the same point.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL PRE-FILLED SALINE SYRINGE
Type of Device
SALINE VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14598596
MDR Text Key294522078
Report Number1911916-2022-00287
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number306546
Device Catalogue Number306546
Device Lot Number306546
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/06/2022
Initial Date FDA Received06/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-