Model Number 37800 |
Device Problems
High impedance (1291); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Vomiting (2144); Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/03/2022 |
Event Type
malfunction
|
Event Description
|
Information was received from a healthcare provider regarding a patient who was implanted with an implantable neurostimulator (ins) for gastric stimulation.It was reported that patients symptoms of nausea and vomiting came back after the device was replaced on (b)(6) 2022. tried to change polarity by turning on the device positive, turning lead 3 negative, turning off lead 2 which had a impedance >20,000 pt is in the process of being scheduled for the device to be evaluated at abbott (b)(4), date of the procedure has not been established.
|
|
Manufacturer Narrative
|
Product id: 4351-35, serial# (b)(4), implanted: (b)(6) 2016, product type: lead.Product id: 4351-35, serial# (b)(4), implanted: (b)(6) 2016, product type: lead.Other relevant device(s) are: product id: 4351-35, serial/lot #: (b)(4), ubd: 11-jul-2018, udi#: (b)(4); product id: 4351-35, serial/lot #: (b)(4), ubd: 11-jul-2018, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the manufacturing representative.It was reported that the cause of the high impedance was unknown, will be identified when dr.Sen chong takes the patient in to investigate.However, the most likely cause or contributing factor to this issue was also unknown, but suspected to be a loose set screw.No further complication was reported at this time.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|