Model Number 989803174201 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Burn(s) (1757)
|
Event Date 05/17/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
A follow up report will be submitted once the investigation is complete.
|
|
Event Description
|
It was reported to philips that the electrode l of the ecg cable began to burn the patient.The device was in use at time of event, a minor burn was reported.
|
|
Manufacturer Narrative
|
The device was returned to philips.Visual inspection of the lead set found that the lead set has burn damaged on wire for the yellow grabber.It cannot be confirmed if the cable is fatigued, or damage was caused by tension on the lead set.The lead set could not be functionally due to hazard of exposed wire.The customer has ordered the replacement part to resolve the issue.The device will be held in the mcs lab for further review if required.
|
|
Event Description
|
It was reported to philips that the electrode l of the ecg cable began to burn the patient.The device was in use at time of event, a minor burn was reported.
|
|
Search Alerts/Recalls
|