Model Number DB-1200-S
Device Problems Delayed Charge Time (2586); Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2022
Event Type  Injury  
Manufacturer Narrative
Date of event: approximated based on the date the manufacturer became aware of the event.
 
Event Description
It was reported that the patient experienced difficulty charging their deep brain stimulation (dbs) implantable pulse generator (ipg).Charging ability did not improve despite additional training and education.The physician assessed the patients weight, in addition to the implant being implanted under the muscle and deeper than the recommended 2-centimeter depth was the cause of the difficulty charging.The patient underwent a procedure where the ipg replaced.The patient did well post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE GEVIA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key14627378
MDR Text Key294384331
Report Number3006630150-2022-02743
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729984443
UDI-Public08714729984443
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/06/2022
Device Model NumberDB-1200-S
Device Catalogue NumberDB-1200-S
Device Lot Number741536
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/17/2022
Initial Date FDA Received06/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/07/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age22 YR
Patient SexFemale
-
-