MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD RESPIRATORY HUMIDIFIER; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD RESPIRATORY HUMIDIFIER; BTT Back to Search Results
Model Number MR850
Device Problems No Audible Alarm (1019); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/03/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint mr850 respiratory humidifier is currently en-route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Event Description
A distributor in south dakota reported that the audible alarm of a mr850 respiratory humidifier was not functioning.There was no patient involvement.
 
Event Description
A distributor in south dakota reported that the audible alarm of a mr850 respiratory humidifier was not functioning.There was no patient involvement.
 
Manufacturer Narrative
Ps396902 method: the pcb of the complaint (b)(4) respiratory humidifier was returned to fisher & paykel healthcare in new zealand for investigation, where it was performance tested and electrically evaluated.Results: performance testing revealed that the pcb did not cause any errors or alarms.Electrical resistance testing revealed the speaker was functioning as intended and the impedence measured was within an acceptable range.Conclusion: we are unable to determine the cause of the reported event as no fault was found with the returned pcb.Our mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the (b)(4) heater base.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The (b)(4) is equipped with visual alarm indicators in addition to the audible alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRATORY HUMIDIFIER
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite #300
irvine, CA 92618
9494534000
MDR Report Key14627658
MDR Text Key295614949
Report Number9611451-2022-00519
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR850
Device Catalogue NumberMR850
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/11/2022
Initial Date FDA Received06/07/2022
Supplement Dates Manufacturer Received07/19/2022
Supplement Dates FDA Received08/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/28/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-