• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problem Mechanical Problem (1384)
Patient Problems Hypoventilation (1916); Low Oxygen Saturation (2477)
Event Date 05/03/2022
Event Type  Injury  
Manufacturer Narrative
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that a patient with aml (acute myeloid leukemia) neutropenic sepsis and pulmonary edema and needing 100% oxygen but managing to clear co2 (carbon dioxide) was put on the 3100a.However, when on the oscillator, ph (potential of hydrogen) decreased from 7.4 to 6.8, and c02 increased from 6 to 21.End user performed hand ventilation and returned the patient back on the 3100a with max map and amplitude with no improvement.Filter helped slightly.The patient was then transferred to hfov b which managed well for 24 hours until the blue filter became really wet and could not maintain map (mean airway pressure).As desaturation began, the patient was transferred to another oscillator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key14628057
MDR Text Key293615402
Report Number2021710-2022-16152
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446005578
UDI-Public(01)10846446005578
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number23689-401
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/11/2022
Initial Date FDA Received06/07/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age6 YR
Patient SexMale
Patient Weight24 KG
-
-