NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MC2-5090S
Device Problems Mechanical Problem (1384); Naturally Worn (2988); Scratched Material (3020)
Patient Problem Failure of Implant (1924)
Event Date 07/08/2021
Event Type  Injury  
Manufacturer Narrative
The device has not been returned for evaluation.The device evaluation was performed by the london implant retrieval center (lirc).The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event Description
An investigation report from an implant retrieval center was received and reported that a revision procedure was performed due to the rod failing to distract.During examination of the rod, the retrieval center found the rod unable to distract.Additionally, discoloration, wear, scratches, and other surface damage was noted.There was no further reported patient impact and no patient or user injury was reported.No additional information is available.Report 1 of 2.
 
Event Description
No additional information has been provided.
 
Manufacturer Narrative
Corrected data: d4.Additional data: h6, h10.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key14629745
MDR Text Key293526523
Report Number3006179046-2022-00198
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00887517834041
UDI-Public887517834041
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMC2-5090S
Device Lot Number7091601AAA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/10/2022
Initial Date FDA Received06/08/2022
Supplement Dates Manufacturer Received06/27/2022
Supplement Dates FDA Received06/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age10 YR
Patient SexFemale
Patient Weight39 KG
-
-