Model Number A209 |
Device Problems
Signal Artifact/Noise (1036); Incorrect, Inadequate or Imprecise Result or Readings (1535); Failure to Convert Rhythm (1540); Failure to Sense (1559); Low impedance (2285)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/24/2022 |
Event Type
Injury
|
Event Description
|
It was reported that at a routine subcutaneous implantable defibrillator (s-icd) device replacement procedure for normal battery depletion, this device was implanted and with the automatic screening tool (ast), was programmed to the same sensing vector as the former device.However, when the physician performed an induction test, the shock was unable to convert the induced ventricular fibrillation (vf).Additionally, the real time internal device electrocardiogram (egm) displayed a flat line showing a lack of sensing.This was followed by high amplitude non-cardiac noisy signals.Because the patient was still in vf, the patient was externally shocked to convert their arrhythmia.The testing result for the induced shock revealed an impedance measurement of 1 ohm.The device was able to interrogated via telemetry and the physician noted an unspecified red alert message.Because there were no additional replacement devices available, the physician turned off the shock therapy and the patient was hospitalized.A surgical replacement procedure is anticipated, but has not yet been performed.At this time, this s-icd device remains implanted and no additional adverse effects were reported.
|
|
Event Description
|
It was reported that at a routine subcutaneous implantable defibrillator (s-icd) device replacement procedure for normal battery depletion, this device was implanted and with the automatic screening tool (ast), was programmed to the same sensing vector as the former device.However, when the physician performed an induction test, the shock was unable to convert the induced ventricular fibrillation (vf).Additionally, the real time internal device electrocardiogram (egm) displayed a flat line showing a lack of sensing.This was followed by high amplitude non-cardiac noisy signals.Because the patient was still in vf, the patient was externally shocked to convert their arrhythmia.The testing result for the induced shock revealed an impedance measurement of 1 ohm.The device was able to interrogated via telemetry and the physician noted an unspecified red alert message.Because there were no additional replacement devices available, the physician turned off the shock therapy and the patient was hospitalized.A few days later, the physician surgically explanted and replaced the entire s-icd system to resolve the event.No additional adverse effects were reported.The system is expected to be returned for analysis, but has not yet been received.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a visual inspection of the device noted an arc mark on the titanium case.Arcing damage like that seen with this device is consistent with an attempt to deliver therapy through a shorted lead system, which may cause internal damage to the circuitry.Based on inspection and analysis of this device, evidence indicates the device was damaged after delivering a shock through a shorted defibrillation electrode.
|
|
Event Description
|
It was reported that at a routine subcutaneous implantable defibrillator (s-icd) device replacement procedure for normal battery depletion, this device was implanted and with the automatic screening tool (ast), was programmed to the same sensing vector as the former device.However, when the physician performed an induction test, the shock was unable to convert the induced ventricular fibrillation (vf).Additionally, the real time internal device electrocardiogram (egm) displayed a flat line showing a lack of sensing.This was followed by high amplitude non-cardiac noisy signals.Because the patient was still in vf, the patient was externally shocked to convert their arrhythmia.The testing result for the induced shock revealed an impedance measurement of 1 ohm.The device was able to interrogated via telemetry and the physician noted an unspecified red alert message.Because there were no additional replacement devices available, the physician turned off the shock therapy and the patient was hospitalized.A few days later, the physician surgically explanted and replaced the entire s-icd system to resolve the event.No additional adverse effects were reported.The system was received for analysis.
|
|
Search Alerts/Recalls
|
|