C.R. BARD, INC. (BASD) -3006260740 POWER PORT MRI ISP, 8 FR CHRONOFLEX, INT WO SP, ATTACH SL; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 8808061 |
Device Problems
Fracture (1260); Dull, Blunt (2407); Deformation Due to Compressive Stress (2889)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The catalog number identified has not been cleared in the us, but is similar to the powerport implantable port products that are cleared in the us.The pro code and 510k number for the powerport implantable port products is identified.As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 06/2023).
|
|
Event Description
|
It was reported that during a port placement procedure, the port sheath tip was allegedly fractured.The procedure was completed using another device.There was no reported patient injury.
|
|
Event Description
|
It was reported that during a port placement procedure, the port sheath tip was allegedly fractured.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport implantable port products that are cleared in the us.The pro code and 510k number for the powerport implantable port products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one powerport mri isp implantable port kit was returned for evaluation.Functional, gross visual and microscopic visual evaluations were performed.The investigation is confirmed for the reported fracture issue and the identified deformation and blunt issues, as the 8.0fr introducer peel-apart sheath and vessel dilator was noted to be bend and fractured.The distal tip of the vessel dilator was noted to be blunt.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 06/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
|