• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWER PORT MRI ISP, 8 FR CHRONOFLEX, INT WO SP, ATTACH SL; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWER PORT MRI ISP, 8 FR CHRONOFLEX, INT WO SP, ATTACH SL; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 8808061
Device Problems Fracture (1260); Dull, Blunt (2407); Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/06/2022
Event Type  malfunction  
Manufacturer Narrative
The catalog number identified has not been cleared in the us, but is similar to the powerport implantable port products that are cleared in the us.The pro code and 510k number for the powerport implantable port products is identified.As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 06/2023).
 
Event Description
It was reported that during a port placement procedure, the port sheath tip was allegedly fractured.The procedure was completed using another device.There was no reported patient injury.
 
Event Description
It was reported that during a port placement procedure, the port sheath tip was allegedly fractured.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport implantable port products that are cleared in the us.The pro code and 510k number for the powerport implantable port products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one powerport mri isp implantable port kit was returned for evaluation.Functional, gross visual and microscopic visual evaluations were performed.The investigation is confirmed for the reported fracture issue and the identified deformation and blunt issues, as the 8.0fr introducer peel-apart sheath and vessel dilator was noted to be bend and fractured.The distal tip of the vessel dilator was noted to be blunt.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 06/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER PORT MRI ISP, 8 FR CHRONOFLEX, INT WO SP, ATTACH SL
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14640939
MDR Text Key293611168
Report Number3006260740-2022-02136
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741098987
UDI-Public(01)00801741098987
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8808061
Device Lot NumberREFZ1358
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/22/2022
Initial Date FDA Received06/09/2022
Supplement Dates Manufacturer Received07/26/2022
Supplement Dates FDA Received07/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-