• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 48-50-00
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/12/2022
Event Type  malfunction  
Event Description
Livanova received report that during priming on the 4 displays of the s5 system panel all the symbols of the sensors did not light up.The symbol with the triangle was displayed and all the sensors were not working.Reportedly, the monitor and associated function of s5 pump did not work.There was no patient involvement.
 
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the s5 system.The incident occurred in (b)(6).A livanova (b)(4) was dispatched to the facility and replaced the dc/dc module of the e/p pack.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
Analysis of service history records revealed that the dc/dc module was also replaced in (b)(6) 2021.However, the dhr review highlighted that the lot whose claimed product belong to was released as conform according to specifications.The root cause of the reported event is a can interruption due to a premature failure of dc/dc module of the ep-pack.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key14641030
MDR Text Key293617537
Report Number9611109-2022-00296
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number48-50-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/12/2022
Initial Date FDA Received06/09/2022
Supplement Dates Manufacturer Received07/12/2022
Supplement Dates FDA Received08/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/22/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-