Model Number 48-50-00 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/12/2022 |
Event Type
malfunction
|
Event Description
|
Livanova received report that during priming on the 4 displays of the s5 system panel all the symbols of the sensors did not light up.The symbol with the triangle was displayed and all the sensors were not working.Reportedly, the monitor and associated function of s5 pump did not work.There was no patient involvement.
|
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the s5 system.The incident occurred in (b)(6).A livanova (b)(4) was dispatched to the facility and replaced the dc/dc module of the e/p pack.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
Analysis of service history records revealed that the dc/dc module was also replaced in (b)(6) 2021.However, the dhr review highlighted that the lot whose claimed product belong to was released as conform according to specifications.The root cause of the reported event is a can interruption due to a premature failure of dc/dc module of the ep-pack.
|
|