• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM Back to Search Results
Model Number 3810002
Device Problems Mechanical Problem (1384); Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/16/2022
Event Type  malfunction  
Event Description
The customer reported that the iv pole drops for no reason.There was not a patient involved at the time the iv pole dropped, therefore no patient information is reasonably known at the time of the event.
 
Manufacturer Narrative
Investigation: the device was inspected by the terumo bct technician.The iv pole adjustment knob was adjusted at the customer site.Investigation is in process.A follow up report will be provided.
 
Manufacturer Narrative
This report is being filed to provide additional information.Investigation: the device was inspected by the terumo bct technician.The iv pole adjustment knob was adjusted at the customer site.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no other problems identified related to the reported condition.Root cause: a root cause assessment was performed for this complaint.The root cause was determined to be related to the need for an iv pole button adjustment.
 
Event Description
The customer reported that the iv pole drops for no reason.There was not a patient involved at the time the iv pole dropped, therefore no patient information is reasonably known at the time of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key14645477
MDR Text Key301574247
Report Number1722028-2022-00193
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583810006
UDI-Public05020583810006
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3810002
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/16/2022
Initial Date FDA Received06/09/2022
Supplement Dates Manufacturer Received07/29/2022
Supplement Dates FDA Received08/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-