BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4483 |
Device Problem
Failure to Disconnect (2541)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/12/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was received for evaluation.Functional testing including leak, clear passage, clamp function, and torque engagement were performed with no issues noted.A connection and disconnection of the female connector with an in-lab connector was performed with no issues noted.The reported connection issue with the female connector was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a minicap transfer set had a connection issue.It was further described as "difficulty removing the transfer set from the manual solution bag and a clamp was used to disconnect the transfer set".This occurred during use of the device for peritoneal dialysis (pd) therapy.To resolve the issue, the transfer set was replaced.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|