SMITH & NEPHEW, INC. GII PS HI FLEX ISRT TR S3-4 10; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 71433453
Device Problem Flaked (1246)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/02/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that genesis ii posterior stabilized high flexion insert trials from size 1-2 10mm to size 7-8 12mm are starting to flake and chip.As this was noticed upon inspection, there was not patient involvement.
 
Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Corrected data: b5, g2.
 
Event Description
It was reported that a gii ps hi flex isrt tr s3-4 10 is starting to flake and chip.As this was noticed upon inspection, there was not patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HI FLEX ISRT TR S3-4 10
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14671966
MDR Text Key301125939
Report Number1020279-2022-02962
Device Sequence Number1
Product Code LXH
UDI-Device Identifier00885556480373
UDI-Public00885556480373
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number71433453
Device Catalogue Number71433453
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/02/2022
Initial Date FDA Received06/12/2022
Supplement Dates Manufacturer Received07/21/2022
Supplement Dates FDA Received07/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-