Reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.K151165.Analysis and results: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have only received some pictures showing a suture with needle detached from the thread.However, without any closed samples a proper analysis cannot be performed.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.Final conclusion: in spite of receiving some pictures showing a defective sample, without closed samples a suitable analysis cannot be performed.Nevertheless, we take note of this incidence and if any closed sample is received in the future, we will re-open the case and analyse it.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
|