• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN SURGICAL, S.A. SUPRAMID BLACK 3/0 (2) 45CM DS19; OTHER SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN SURGICAL, S.A. SUPRAMID BLACK 3/0 (2) 45CM DS19; OTHER SUTURE Back to Search Results
Model Number C0712213
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/19/2022
Event Type  malfunction  
Event Description
It was reported an issue with supramid suture.The customer (veterinarian) reported that the needle comes off the thread immediately every time the package is opened.No further information has been provided.
 
Manufacturer Narrative
Reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.K151165.Analysis and results: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have only received some pictures showing a suture with needle detached from the thread.However, without any closed samples a proper analysis cannot be performed.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.Final conclusion: in spite of receiving some pictures showing a defective sample, without closed samples a suitable analysis cannot be performed.Nevertheless, we take note of this incidence and if any closed sample is received in the future, we will re-open the case and analyse it.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPRAMID BLACK 3/0 (2) 45CM DS19
Type of Device
OTHER SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key14673257
MDR Text Key295501621
Report Number3003639970-2022-00250
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0712213
Device Catalogue NumberC0712213
Device Lot Number621324
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/19/2022
Initial Date FDA Received06/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-