• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUSELECT SUCTIONAID TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUSELECT SUCTIONAID TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/860/090CZ
Device Problem Disconnection (1171)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2022
Event Type  malfunction  
Event Description
It was reported that eight days after using this tracheostomy tube from regular replacement, the customer noticed that the pilot balloon detached.The patient was breathing spontaneously and did not pull the pilot balloon.No patient injury was reported.
 
Manufacturer Narrative
A sample was received to perform an investigation.A visual inspection of the returned device found that the inflation line was detached from the pilot balloon.The reported problem was confirmed from visual inspection.A non conformance report (ncr) was initiated to further investigate this supplied item.At this time, the root cause of the failure was unable to be determined.A device history record (dhr) review was performed and showed that there were no observations recorded during manufacture to suggest an issue of this nature occurred with this product lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUSELECT SUCTIONAID TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-1-1 akasaka,
minneapolis, MN 55442
MDR Report Key14674831
MDR Text Key295318435
Report Number3012307300-2022-11662
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076172
UDI-Public15019517076172
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/090CZ
Device Catalogue Number101/860/090CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/16/2022
Initial Date FDA Received06/13/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexMale
-
-