• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Model Number N/A
Device Problems Break (1069); Unstable (1667); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/20/2022
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report the cable break.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.The steerable guide catheter (sgc) was inserted into the left atrium and the clip delivery system (cds) was advanced in the patient.The plus/minus knob was turned about a quarter when a click sound was heard, and the knob no longer worked.The cds and sgc were removed from the patient.Troubleshooting was performed outside the patient and the knob showed no resistance as it was continuously spinning.The physician suspected a cable break.A new sgc was selected and the same cds was used to complete the procedure reducing the final mr to a grade of 2.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
All available information was investigated and the reported cable break was confirmed via returned device analysis.The reported noise could not be replicated in a testing environment as it was likely a symptom of the cable break.The reported unstable/loose knob was not confirmed.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation determined the reported noise and unstable/loose knob appear to be related to the reported broken cable.However, a cause for how the cable broke cannot be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14675151
MDR Text Key293845974
Report Number2024168-2022-06313
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/09/2023
Device Model NumberN/A
Device Catalogue NumberSGC0702
Device Lot Number20106R115
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/20/2022
Initial Date FDA Received06/13/2022
Supplement Dates Manufacturer Received07/18/2022
Supplement Dates FDA Received07/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CLIP DELIVERY SYSTEM
-
-