• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARCOS PROXIMAL BODY INSERTER; INSTRUMENT, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARCOS PROXIMAL BODY INSERTER; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Device Problems Difficult to Advance (2920); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Visual examination of the provided pictures identified the shaft has wear lines, and there a spot of minor damage/deformation to the threads.No rolled threads are visible.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the proximal body inserter was cross threaded.There was no impact to patient and no adverse consequences were reported.During investigation, the threads were observed to be damaged.No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOS PROXIMAL BODY INSERTER
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14681089
MDR Text Key293925017
Report Number0001825034-2022-01388
Device Sequence Number1
Product Code KWA
UDI-Device Identifier00880304490093
UDI-Public(01)00880304490093(11)210924(10)75922105
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090757
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number31-301000
Device Lot Number75922105
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/08/2022
Initial Date FDA Received06/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/24/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-