• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART XL+ DEFIBRILLATOR/MONITOR; DFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART XL+ DEFIBRILLATOR/MONITOR; DFIBRILLATOR Back to Search Results
Model Number 861290
Device Problems Device Alarm System (1012); Failure of Device to Self-Test (2937)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2022
Event Type  malfunction  
Event Description
It was reported that rfu shows "x".There was reportedly no patient involvement.Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Manufacturer Narrative
The manufacturer's authorized field service engineer (fse) evaluated the device and confirmed the reported problem.Upon conclusion of the evaluation, it was determined that the battery, therapy pca and processor pca required replacement.We are unable to determine the root cause of the event as multiple parts were required for replacement.The device remains at the customer site and no further evaluation is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART XL+ DEFIBRILLATOR/MONITOR
Type of Device
DFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
daniel derochers
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key14683915
MDR Text Key293926403
Report Number3030677-2022-02987
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838023680
UDI-Public00884838023680
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number861290
Device Catalogue Number861290
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/09/2022
Initial Date FDA Received06/14/2022
Supplement Dates Manufacturer Received06/09/2022
Supplement Dates FDA Received08/18/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-