RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/02/2022
Event Type  malfunction  
Event Description
The customer reported the ventilator alarms for a high leak rate.No patient involvement.The remote service engineer advised customer to perform a complete performance verification test (pvt) to determine if unit is functioning as intended.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
 
Manufacturer Narrative
Reporting institution name: (b)(6) hospital.
 
Manufacturer Narrative
The customer reported he performed a system and high pressure leak tests multiple times with no issues.He then completed a full performance verification testing of unit with calibrated verified equipment and found zero errors with good results through pressure and flow tests.He also ran with artificial lung as well.No errors were found in service log either.Customer put the unit back into service and have had no news of any further issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
kimberly shelly
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key14686487
MDR Text Key301723065
Report Number2031642-2022-01607
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/02/2022
Initial Date FDA Received06/14/2022
Supplement Dates Manufacturer Received06/13/2022
Supplement Dates FDA Received07/08/2022
Date Device Manufactured10/08/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-