• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC. ORBERA¿ INTRAGASTRIC BALLOON SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC. ORBERA¿ INTRAGASTRIC BALLOON SYSTEM Back to Search Results
Model Number B-4800
Device Problem Insufficient Information (3190)
Patient Problem Hemorrhage/Bleeding (1888)
Event Type  malfunction  
Manufacturer Narrative
Initial medwatch submitted to the fda on 15/jun/2022.A review of the device labeling notes the following: the current orbera¿ intragastric balloon system directions for use (dfu) addresses the known and anticipated potential events of ulcer; hemorrhage; and vessel damage / bleeding as follows: the igb is composed of a soft silicone elastomer and is easily damaged by instruments or sharp objects.The igb must be handled only with gloved hands and with the instruments recommended in this document.The physiological response of the patient to the presence of orbera may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response.Each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms.Patients with an igb that present with severe abdominal pain that have a negative endoscopy and x-ray may additionally require a ct scan to definitively rule out a perforation.Patients must be advised that the igb is intended to be placed for 6 months maximally, at which point removal is required.Longer periods of igb placement increase the risk of igb deflation (a reduction in size of the device due to loss of saline) which can lead to intestinal obstruction and risk for death.The risk of these events is also significantly higher when igbs are filled to a larger volume than indicated (greater than 700cc).It is important to discuss all possible complications and adverse events with your patient.Complications that may result from the use of this product include the risks associated with the medications and methods utilized in the endoscopic procedure, the risks associated with any endoscopic procedure, the risks associated with the orbera intragastric balloon specifically, and the risks associated with the patient's degree of intolerance to a foreign object placed in the stomach.Possible complications: gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon.Continuing nausea and vomiting.This could result from direct irritation of the lining of the stomach or as a result of the balloon blocking the outlet of the stomach.It is even theoretically possible that the balloon could prevent vomiting (not nausea or retching) by blocking the inlet to the stomach from the esophagus.Abdominal or back pain, either steady or cyclic gastroesophageal reflux.Insufficient or no weight loss.Additional information: we are waiting to hear back from doctor to see if it a ulcer.Will update on supplement report.The device has not been returned for analysis.The investigator is waiting until the lot number of the device is known to determine whether a device history record (dhr) review is or is not required for this complaint.
 
Event Description
After 72 hours of balloon implant, patient vomiting blood after taken asprin.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBERA¿ INTRAGASTRIC BALLOON SYSTEM
Type of Device
INTRAGASTRIC BALLOON SYSTEM
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX 78746
Manufacturer (Section G)
APOLLO ENDOSURGERY COSTA RICA, SRL
coyol free zone
building b 13.3
alajuela, cs CRI
CS   CRI
Manufacturer Contact
david hooper
1120 s. captail of texas hwy
bldg 1, ste 300
austin, TX 78746
5128523757
MDR Report Key14695028
MDR Text Key295810407
Report Number3006722112-2022-00053
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberB-4800
Device Catalogue NumberB-4800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/18/2022
Initial Date FDA Received06/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-