Model Number 37800
Device Problems Impedance Problem (2950); Material Deformation (2976)
Patient Problems Nausea (1970); Pain (1994)
Event Date 05/23/2022
Event Type  Injury  
Event Description
Information was received from a healthcare provider regarding a patient who was implanted with an implantable neurostimulator (ins) for gastrointestinal/pelvic floor.It was reported that patient reported to the hospital with increased nausea and pain.Impedance was out of range 20,000.A x-ray showed the leads were in knots within the abdomen.The hcp plans on removing both the leads and battery and replacing them.
 
Manufacturer Narrative
Concomitant medical products: product id: 4351-35, serial# (b)(4), implanted: (b)(6) 2019, explanted: product type: lead.Product id: 4351-35, serial# (b)(4), implanted: (b)(6) 2019, explanted: product type: lead.Other relevant device(s) are: product id: 4351-35, serial/lot #: (b)(4), ubd: 29-may-2021, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a manufacturer representative (rep).The rep reported that patient had a new battery on july 8th.
 
Manufacturer Narrative
Continuation of d10: product id neu_unknown lot# serial# unknown implanted: explanted: product type unknown product id 4351-35 lot# serial# (b)(6), implanted: (b)(6)2019, explanted: product type lead product id 4351-35 lot# serial# (b)(6), implanted: (b)(6)2019, explanted: product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14732961
MDR Text Key294264283
Report Number3004209178-2022-07773
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169614246
UDI-Public00643169614246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2020
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/15/2022
Initial Date FDA Received06/17/2022
Supplement Dates Manufacturer Received09/27/2022
Supplement Dates FDA Received09/29/2022
Date Device Manufactured04/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient SexFemale
-
-