The customer/user reported to olympus she had a chemical exposure after a suspected ventilation issue with the room the subject device was located.The customer advised after 2 days red blotches were present on her face, welts were present in the corner of her eyes, and the inside of her nose was dried out.The user works in the same room as the reprocessor for her entire shift 8:00 am to 4:30 pm.She wears gloves at a minimum when handling any chemicals.User advised she reported incident to locations management.The user advised her location facilities whom tested/adjusted the location hvac system and issue has been resolved.Contact advised that she has since recovered from symptoms.Olympus field service engineer (fse) spoke with the customer and verified the splash guard was in place, the gas filters were in place and no leaks were present.Fse went onsite and confirmed the customer's units are in good working condition.No issues found.
|