INTUITIVE SURGICAL, INC NONE; ENDOSCOPE PLUS Back to Search Results
Model Number 470057-08
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/06/2022
Event Type  malfunction  
Event Description
It was reported that during a da vinci assisted surgical procedure, the endoscope plus 30 degree was observed with control that was displayed upside down during the operation.The procedure was completed with no reported injury.On may 20, 2021, intuitive surgical, inc.(isi) contacted the site to obtain additional information regarding this event.However, no additional details have been received as of the date of this report.
 
Manufacturer Narrative
Isi received the endoscope involved with this complaint and completed the device evaluation.Failure analysis (fa) investigation confirmed the customer reported complaint.A damaged or friction issue was noted on the camera instrument adapter.Fa also observed discoloration on the housing of the endoscope.The root cause was attributed to mishandling/misuse.No photo or video was provided by the site for review.A review of the device logs for the 30 degree endoscope (part# 470057-08 | lot/serial# (b)(4)) associated with this event has been performed.Per this review of the logs, the 30 degree endoscope was last used on (b)(6) 2022 via system serial# (b)(4).There were 1942 uses remaining after this last usage.This complaint is considered as reportable due to the following conclusion: it was alleged that the endoscope moved with unintuitive motion with no evidence or claim of user mishandling or misuse.Poor camera control could result in unintuitive motion and subsequent tissue damage.At this time, the root cause of the failure is unknown.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONE
Type of Device
ENDOSCOPE PLUS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key14740084
MDR Text Key301930910
Report Number2955842-2022-12244
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00886874116555
UDI-Public(01)00886874116555
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K191736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470057-08
Device Catalogue Number470057
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/19/2022
Initial Date FDA Received06/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/02/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-