BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4483 |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/26/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the patient connector of a minicap transfer set was unable to connect tightly to the titanium adapter.This was observed when connecting the two devices, before peritoneal dialysis therapy.The titanium adapter remained in use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional tests which included leak, clear passage, and clamp function tests were performed with no issues noted.The reported condition was not verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|