MAUDE Adverse Event Report: MEDLINE INDUSTRIES LP NEPTUNE HEATED HUMIDIFIER,CONCHASMART

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE INDUSTRIES LP NEPTUNE HEATED HUMIDIFIER,CONCHASMART Back to Search Results
Catalog Number HUD42500
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Partial thickness (Second Degree) Burn (2694)
Event Date 05/29/2022
Event Type  Injury  
Manufacturer Narrative
It was reported by the facility that a newborn was placed on heated high flo oxygen via nc at 2 lpm 30% temp 31.9 c.The respiratory therapist was called to the nursery by the nurse, who stated water was in the tubing of the high flow nasal cannula.It was noted that the nasal cannula was not in the baby's nose; it was hanging over the side of the safety rail at the head of the bed.The nurse had taken off the nasal cannula when the respiratory tech was called in because of water in the tubing.When the respiratory tech began to disconnect the equipment from the end of the tubing, closest to the device, water was released from the device, which was very hot.The tubing was disconnected at the heating element.That is when water was coming out of the tubing onto the baby.The doctor examined the baby, which appeared to have redness on the stomach and chin areas.The doctor ordered to change o2 to a non-heated nasal cannula.After the situation was resolved, all of the equipment associated with the water overheating was sequestered.The sample was not returned to the manufacturer for evaluation it was returned to a biomed company utilized by the hospital.There was no impact to the patient's stability or oxygen level as a result of the incident however the newborn did experience second degree burns on the right hand, chest, right and left shoulder, chin and neck areas from water in the circuit.Wound care nurses began following and treating the newborn on 5/29/2022.On 6/6/2022 the wound care nurse reported that all areas of open skin were healed.Due to the reported incident and in an abundance of caution, this medwatch is being filed.If any further relevant information is identified or obtained, a supplemental medwatch will be submitted.
 
Event Description
It was reported that a newborn sustained second degree burns after being exposed to hot water from the device that was in the tubing hanging on the side of the bed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEPTUNE HEATED HUMIDIFIER,CONCHASMART
Type of Device
NEPTUNE HEATED HUMIDIFIER,CONCHASMART
Manufacturer (Section D)
MEDLINE INDUSTRIES LP
three lakes drive
northfield IL 60093
Manufacturer Contact
karen trutsch
three lakes drive
northfield, IL 60093
MDR Report Key14753518
MDR Text Key294397268
Report Number1417592-2022-00083
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHUD42500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/06/2022
Initial Date FDA Received06/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age1 DA
Patient SexMale
Patient Weight3 KG
-
-