• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN STEM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN STEM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Unspecified Infection (1930); Insufficient Information (4580)
Event Date 06/08/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Report source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for the investigation as the device location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-01475, 0001825034-2022-01476, 0001825034-2022-01477.
 
Event Description
It was reported that patient underwent a hip revision after an unknown amount of time post implantation due to unknown reasons.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that patient underwent a hip revision after an unknown amount of time post implantation due to infection.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6: component code: mechanical (g04) ¿ stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Part and lot identification are necessary for review of device history records, neither were provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN STEM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14776958
MDR Text Key294929084
Report Number0001825034-2022-01478
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/15/2022
Initial Date FDA Received06/22/2022
Supplement Dates Manufacturer Received06/23/2022
09/12/2022
Supplement Dates FDA Received07/18/2022
09/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-