Model Number AB2000
Device Problem Failure to Prime (1492)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/24/2022
Event Type  malfunction  
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that during the aquablation procedure the aquabeam handpiece was unable to prime.As a result, the aquabeam handpiece was replaced but the same issue occurred.A third handpiece unit was used, and the procedure was then continued through successful completion.The reported event caused a procedural delay of over 20 minutes.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Manufacturer Narrative
H.10 additional manufacturer narrative: the aquabeam handpiece was returned for investigation of this event.The handpiece was tested on the qa test system.The cpu light was green indicating that the handpiece was detected and connected.No issues were observed after three docking cycles.The handpiece was also able to prime at 50% and 100% pump power.Additionally, a fully simulated aquablation session was performed successfully.The handpiece was then deconstructed and viewed under magnification.No signs of fluid ingress or any damage were observed.The handpiece functioned as intended and no issues were found.A review of the device history record (dhr) ab2000-b/serial number (b)(6) and associated aquabeam handpiece/ lot number 21c00783r1 was conducted, which confirmed that there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system met all design and manufacturing specifications when released for distribution.The aquabeam robotic system user manual, um0104 rev.F, states the following: 11.2.7 sterile: motorpack draping and docking with the aquabeam handpiece ensure the high-pressure tubing is centered and seated in the lower box.Dock the motorpack to the aquabeam handpiece: verify aquabeam handpiece nozzle position (i.E.The movement of the blue led) homes to the base of the aquabeam handpiece (fully proximal).Apply sterile tape over the connection between aquabeam handpiece and motorpack to seal it.-keep the motorpack and the aquabeam handpiece assembly with the scope clamp assembly in a secure and sterile environment.Note: verify the motorpack is securely engaged to the aquabeam handpiece by attempting to pull the aquabeam handpiece off of the motorpack.The aquabeam handpiece should remain connected to the motorpack if properly engaged.A similar complaint review for a 12-month period was conducted, which confirmed five (5) similar events have been reported to procept.The in-house investigation was unable to reproduce the reported failure mode; the returned device functioned as intended.Complaint data is monitored and trended as part of procept's quality management system; shall a trend arise for this failure mode, then further actions will be considered.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer Contact
doria esquivel
900 island drive
suite 101
redwood city, CA 94065-1494
MDR Report Key14778120
MDR Text Key302030467
Report Number3012977056-2022-00078
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D20190910T
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAB2000
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2022
Initial Date FDA Received06/22/2022
Supplement Dates Manufacturer Received07/20/2022
Supplement Dates FDA Received07/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/10/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-