Model Number 560BCS1 |
Device Problems
Display or Visual Feedback Problem (1184); Smoking (1585)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/14/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during use of a bio-console 560 instrument , it was reported that there was a burnt smell, smoke, black screen when booting.The instrument was replaced to complete the procedure.There were no adverse patient effects because of this issue.Medtronic received additional information stating that the hand crank was required to maintain flow.The hand crank was in use for 3 hours.The issue happen when the bio-console was turned on the first time for this clinical case.The bio-console was being used for an extra-corporeal membrane oxygenation (ecmo) case.The bio-console had been used by the patient for 1 hour when the issue occurred.There was no fluid running through the pump when the event occurred.
|
|
Manufacturer Narrative
|
Product analysis: the reported burnt smell, smoke, black screen when booting was verified during service.The service technician determined that the fault was the power supply, mpu board.The service technician also found that the ui xb display was damaged.The issue will be resolved by replacing the power supply, mpu board, ui xb display.Preventive maintenance was performed per specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Conclusion: the complaint was confirmed for the reported burnt smell, smoke, and black screen when booting.The service technician determined that the fault was the power supply, mpu (assy 560 bio-console syscon) board.The service technician also found that the ui xb display was damaged.The issue was resolved by replacing the power supply 28v, assy 560ui bitsy xb led display and the assy 560 bio-console syscon.Preventive maintenance was performed per specifications.This occurrence is the result of the replacement of a component used in the previous version of the assy 560 bio-console syscon.There were no patient/clinical safety issues reported.Trends for issues with this instrument are reviewed at quarterly quality meetings.Correction to h8 (usage of device): this field has been updated to reuse medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|