• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD10
Device Problem Unintended System Motion (1430)
Patient Problem Pain (1994)
Event Date 06/10/2022
Event Type  Injury  
Event Description
It has been reported to philips that while transferring the patient from the table to the stretcher, the table unexpectedly pivoted.The patient is approximately 100kg and has no legs.Three staff members prevented him from falling and were injured.This report concerns a nurse who injured their back while helping to catch the patient and keep him from falling.It has been reported that the nurse is since wearing a back support belt, which is used to support the lower back (lumbar) region, in particular when lifting is required and helps prevent discomfort.No harm has been reported on the patient.As this incident resulted in three staff members being injured and a patient nearly falling off the table, four complaints have been reported.The other three reports can be found under philips reference numbers (b)(4).Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated the reported problem.A philips service engineer inspected the system on site and identified a deposit of contrast product on the table pivot which resulted in less resistance when in a specific pivot stop position.The deposit was removed, and the system was returned to use in good working order.Corrected data: coding updated based on investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key14786407
MDR Text Key294567003
Report Number3003768277-2022-00295
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/10/2022
Initial Date FDA Received06/23/2022
Supplement Dates Manufacturer Received06/10/2022
Supplement Dates FDA Received01/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-