• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. RAPIDPOINT 500E BLOOD GAS SYSTEM; RP 500E

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. RAPIDPOINT 500E BLOOD GAS SYSTEM; RP 500E Back to Search Results
Catalog Number 11416755
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/16/2022
Event Type  malfunction  
Manufacturer Narrative
Siemens has completed the investigation.Based on the co-ox subcomponent and sample quality checks, aqc performance, lack of any co-ox related calibration drift or diagnostic errors leading up to and on the day of the escalated events, the cooximetry optical subsystem responsible for the measurement of thb on the system appeared stable and functioning as expected.The reported thb values appear valid.The cause of this event is unknown.
 
Event Description
Customer reported that they received multiple discrepant low thb results across multiple days ranging from 2.2 - 5.9 g/dl for the same patient on their rp500e instrument compared to earlier measurements in same device and retesting of different samples on lab instrument.Patient received blood trx (5/21 thb = 11.9 g/dl).There was no reported injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500E BLOOD GAS SYSTEM
Type of Device
RP 500E
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
thomas myers
2 edgewater drive
norwood, MA 02062
MDR Report Key14788824
MDR Text Key302940718
Report Number3002637618-2022-00039
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K192240
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number11416755
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/26/2022
Initial Date FDA Received06/23/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-