Catalog Number 11416755
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2022
Event Type  malfunction  
Event Description
Customer reported that they received a discrepant low thb result for a patient on their rp500e instrument compared to other results of retesting of different samples on the same instrument.There was no reported injury due to this event.
 
Manufacturer Narrative
Siemens has requested instrument log files for further investigation.Customer has provided the files and investigation is underway.The cause of this event is unknown.
 
Manufacturer Narrative
Siemens has completed the investigation.Based on the co-ox subcomponent and sample quality checks, aqc performance, lack of any co-ox related calibration drift or diagnostic errors leading up to and on the day of the escalated events, the cooximetry optical subsystem responsible for the measurement of thb on the system appeared stable and functioning as expected.The reported thb values appear valid.The cause of this event is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500E BLOOD GAS SYSTEM
Type of Device
RP 500E
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING LTD.
northern road
chilton industrial area
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
thomas myers
2 edgewater drive
norwood, MA 02062
MDR Report Key14789269
MDR Text Key302341497
Report Number3002637618-2022-00041
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K192240
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number11416755
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/14/2022
Initial Date FDA Received06/23/2022
Supplement Dates Manufacturer Received06/30/2022
Supplement Dates FDA Received07/20/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-