HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; ULTRASOUND VIDEO BRONCHOSCOPE Back to Search Results
Model Number EB19-J10U
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/25/2022
Event Type  malfunction  
Event Description
The time of event is unknown.There was no report of patient harm.Air/water tube clogged.
 
Manufacturer Narrative
We checked the returned unit and confirmed that the balloon feeder/return tube clogged.Based on the result, we concluded that it was caused due to the insufficient reprocessing at the facility on the balloon feeder/return tube.In addition, we confirmed that the angle wire play; however, it is not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0630(air/water & jet water channels)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
ULTRASOUND VIDEO BRONCHOSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key14796593
MDR Text Key303236372
Report Number9610877-2022-55014
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
K183516
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEB19-J10U
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2022
Initial Date Manufacturer Received 05/25/2022
Initial Date FDA Received06/24/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-