• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWER PORT ISP M.R.I., 8FR GROSHONG; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWER PORT ISP M.R.I., 8FR GROSHONG; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 4808570J
Device Problems Difficult to Insert (1316); Peeled/Delaminated (1454); Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/23/2021
Event Type  malfunction  
Event Description
It was reported that during a port placement procedure via the internal jugular vein, the guidewire allegedly found to be peeled.It was further reported that the device had resistance issue on insertion.There was no reported patient injury.
 
Manufacturer Narrative
The catalog number identified in section d4 has not been cleared in the us but is similar to the power port isp m.R.I., 8fr groshong that are cleared in the us.The pro code and 510 k number for the power port isp m.R.I., 8fr groshong are identified.Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one powerport mri isp implantable port, one flushing connector and one cath-lock attached to a groshong catheter, one introducer needle, one straight non-coring needle, one right-angle non-coring needle, one vein pick, one cath-lock, one syringe, one j-tip guidewire in a guidewire hoops, one 8.0fr introducer peel-apart sheath and vessel dilator, one safety infusion set, one tunneler and one guidewire loaded in a guidewire hoop were returned for evaluation.Gross, microscopic visual and functional testing were performed.The investigation is confirmed for the identified guidewire deformation issue as the guidewire appeared to be bend on the distal tip.However, the investigation is inconclusive for the reported difficulty in insertion issue as the exact circumstances at the time of reported event was unknown and the clinical conditions cannot be replicated to confirm these failures.Furthermore, the investigation is unconfirmed for reported peeled issue as no peeling was noted on the received physical sample.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiration date: 08/2023).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER PORT ISP M.R.I., 8FR GROSHONG
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14797130
MDR Text Key296717272
Report Number3006260740-2022-02421
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741175725
UDI-Public(01)00801741175725
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4808570J
Device Lot NumberREFU1451
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/31/2022
Initial Date FDA Received06/24/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-