MEDTRONIC IRELAND MICRA; INTRODUCER, CATHETER Back to Search Results
Model Number MI2355A
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2022
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the implant procedure, numerous attempts to suck back fluid from the introducer and could not draw back and remove air.Introducer was removed and another one was used successfully. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the introducer was returned and analyzed.The distal seal of the delivery system introducer was torn.Blood was observed on the delivery system introducer flush port valve.Blood was observed on the delivery system introducer flush tube.Blood was observed on the delivery system introducer distal seal.Blood was observed on the delivery system introducer proximal seal.Visual analysis of the introducer indicated damage during use.The analyst noted the introducer was returned with the dilator fully inserted in the sheath of the introducer.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRA
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
EI  
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14797148
MDR Text Key302680878
Report Number9612164-2022-02392
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K132030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/08/2024
Device Model NumberMI2355A
Device Catalogue NumberMI2355A
Device Lot Number00166490
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/21/2022
Initial Date FDA Received06/24/2022
Supplement Dates Manufacturer Received10/11/2022
Supplement Dates FDA Received10/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-