• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. 3.0 RIO® ROBOTIC ARM - MICS; STEREOTAXIC DEVICE, ROBOTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. 3.0 RIO® ROBOTIC ARM - MICS; STEREOTAXIC DEVICE, ROBOTICS Back to Search Results
Model Number 209999
Device Problem Unintended System Motion (1430)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2022
Event Type  malfunction  
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event Description
Mps reported arm dropping out of haptics after motorized alignment.Case completed robotically.
 
Event Description
Mps reported arm dropping out of haptics after motorized alignment.Case completed robotically.
 
Manufacturer Narrative
An event regarding unintended movement involving a mako robotic arm was reported.The event was not confirmed.Method & results: product evaluation and results: the field service engineer reported: problem reproduced? no.Trouble shooting notes: (b)(6) - mps reported arm dropping out of haptics after motorized alignment.Work performed: set up the system and do a successful pre surgery check.Check right and left accuracy - all tests have been nominal.Right side of the robot shown medium numbers of accuracy - new calibrated with great numbers.Work order disposition: system investigation completed successfully as per service manual.All system checks and tests passed, system is ready for use.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate that devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the alleged failure mode was not confirmed.Successfully completed all checks, verifications, and calibrations.System is ready for clinical use.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.H3 other text : not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.0 RIO® ROBOTIC ARM - MICS
Type of Device
STEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
juana hiciano
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14807516
MDR Text Key298718839
Report Number3005985723-2022-00082
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00848486030407
UDI-Public00848486030407
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number209999
Device Catalogue Number209999
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2022
Initial Date FDA Received06/24/2022
Supplement Dates Manufacturer Received08/24/2022
Supplement Dates FDA Received09/19/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-