• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510890
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/06/2022
Event Type  malfunction  
Manufacturer Narrative
The exact date of the event is unknown.The provided event date was chosen as a best estimate based on the date that the manufacturer became aware of the event.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a trapezoid basket sterile pouch was found torn.During unpacking of the shipping box, one of trapezoid basket sterile pouch was found to be torn.This event did not occur during a procedure, did not involve a patient, and occurred outside of an operating room.
 
Manufacturer Narrative
Block h2: additional information: b3 date of event has been updated based on additional information received.Block h6: device code a020504 captures the event tear in device packaging.Block h10: the returned trapezoid basket was analyzed, and a visual evaluation observed that the packaging has a cutting mark from side to side with a tear in the middle.The packaging also has other cutting marks possibly made with a sharp object.The packaging keeps factory seals.The customer provided 3 pictures related to the complaint.The pictures showed the packaging of the device with a tear on it, also, the cardboard box where the damaged device was located.The reported complaint is confirmed.Based on all available information, it is possible that the problem could have happened during transport, or while the device was in storage.Therefore, the most probable root cause is cause traced to transport/storage.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event Description
It was reported to boston scientific corporation that an trapezoid basket sterile pouch was found torn on june 6, 2022.During unpacking of the shipping box, one of trapezoid basket sterile pouch was found to be torn.This event did not occur during a procedure, did not involve a patient, and occurred outside of an operating room.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14836368
MDR Text Key303245134
Report Number3005099803-2022-03466
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296409
UDI-Public08714729296409
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/07/2023
Device Model NumberM00510890
Device Catalogue Number1089
Device Lot Number0029187055
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/07/2022
Initial Date FDA Received06/27/2022
Supplement Dates Manufacturer Received07/06/2022
Supplement Dates FDA Received08/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-