Catalog Number 1857SI22 |
Device Problems
Partial Blockage (1065); Component Misassembled (4004)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
|
Event Description
|
It was reported that the the bladder flush did not flow after the insertion of the foley catheter.After, the catheter was removed, the user notice that the rinsing channel of the catheter was obstructed.The catheter was replaced with a new one.
|
|
Event Description
|
It was reported that the the bladder flush did not flow after the insertion of the foley catheter.After, the catheter was removed, the user notice that the rinsing channel of the catheter was obstructed.The catheter was replaced with a new one.
|
|
Manufacturer Narrative
|
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿irrigation lumen coverage too thin".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.¿ h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
|
Manufacturer Narrative
|
The reported event is confirmed, manufacturing related.The red rubber foley catheter was returned without the original packaging.The irrigation eye was not burned all the way through which would prevent the water from flowing out.A potential root cause for this event could be, "operator error".No manufacturing issues or non-conformances were noted during review of the dhr that could have caused or contributed to the reported event.Based on the results of the investigation no additional action is required at this time.A label/pack review is not required as a review of the label could not have prevented the reported event.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
|
Event Description
|
It was reported that the the bladder flush did not flow after the insertion of the foley catheter.After, the catheter was removed, the user notice that the rinsing channel of the catheter was obstructed.The catheter was replaced with a new one.
|
|