Catalog Number 1857SI22
Device Problems Partial Blockage (1065); Component Misassembled (4004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the the bladder flush did not flow after the insertion of the foley catheter.After, the catheter was removed, the user notice that the rinsing channel of the catheter was obstructed.The catheter was replaced with a new one.
 
Event Description
It was reported that the the bladder flush did not flow after the insertion of the foley catheter.After, the catheter was removed, the user notice that the rinsing channel of the catheter was obstructed.The catheter was replaced with a new one.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿irrigation lumen coverage too thin".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.¿ h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Manufacturer Narrative
The reported event is confirmed, manufacturing related.The red rubber foley catheter was returned without the original packaging.The irrigation eye was not burned all the way through which would prevent the water from flowing out.A potential root cause for this event could be, "operator error".No manufacturing issues or non-conformances were noted during review of the dhr that could have caused or contributed to the reported event.Based on the results of the investigation no additional action is required at this time.A label/pack review is not required as a review of the label could not have prevented the reported event.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that the the bladder flush did not flow after the insertion of the foley catheter.After, the catheter was removed, the user notice that the rinsing channel of the catheter was obstructed.The catheter was replaced with a new one.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14836701
MDR Text Key301638396
Report Number1018233-2022-04929
Device Sequence Number1
Product Code MJC
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/30/2023
Device Catalogue Number1857SI22
Device Lot NumberMCCQ0350
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/09/2022
Initial Date FDA Received06/27/2022
Supplement Dates Manufacturer Received07/21/2022
11/28/2022
Supplement Dates FDA Received08/04/2022
11/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-