CYBERONICS - HOUSTON LEAD MODEL 302 Back to Search Results
Model Number 302-20
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Neck Pain (2433); Fibrosis (3167); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 02/11/2022
Event Type  Injury  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ;defects¿ or ;malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.(b)(4).
 
Event Description
It was reported that the patient had gained weight recently and enlargement of her breast which caused traction in the lead on the left side of the neck and the lead to be more prominent and restricting her neck movement.The patient had imaging done and the vns was confirmed to be intact and all values when checked were within normal limits.The patient requested to have repositioning surgery.During the surgery the surgeon noticed scar tissue and calcification around the lead that was dissected and then the lead was replaced.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
dana sprague
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key14837648
MDR Text Key294870666
Report Number1644487-2022-00770
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/18/2009
Device Model Number302-20
Device Lot Number1534
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 06/02/2022
Initial Date FDA Received06/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/20/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age38 YR
Patient SexFemale
-
-