• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH CRANIAL-SCR PLUSDRIVE Ø1.6 SELF-DRILL L4; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH CRANIAL-SCR PLUSDRIVE Ø1.6 SELF-DRILL L4; PLATE, BONE Back to Search Results
Catalog Number 400.834
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/30/2022
Event Type  malfunction  
Event Description
Device report from synthes reports an event in china as follows: it was reported that the device could not self drill.It was reported that during the surgery, when insert the screw, noted the self drilling screws could not drill.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.This complaint involves one(1) device cranial-scr plusdrive ø1.6 self-drill l4.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device available for evaluation: complainant part is not expected to be returned for manufacturer review/investigation.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d4, d9 h3, h6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the cranial-scr plusdrive ø1.6 self-drill l4 had the edges of the threads and the distal tip worn.The device was received with white organic particles attached along the threads and unknown colored substance on the head.Additionally, it presents nicks and wear on the drive likely from use/engagements of the screwdriver tip, this condition cannot be traced to a manufacturing issue.The device returned is a representative sample and not the complaint device.A dimensional inspection for the cranial-scr plusdrive ø1.6 self-drill l4 was unable to be performed due to post manufacturing damage.A functional test was unable to be performed since the functionality of the screw cannot be proven with tools.The complaint condition was not able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the cranial-scr plusdrive ø1.6 self-drill l4 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H6: a device history record (dhr) review was conducted: manufacturing location: monument, manufacturing date: 05-may-2021, part number: 400.834, ti low profile neuro screw self-drilling 4mm, lot number: 149p092 (non-sterile), lot quantity: (b)(4).Production order traveler met all inspection acceptance criteria.Inspection sheet, inspect dimensional, met all inspection acceptance criteria.Packaging label log (pll) was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: part number: 21015, tialnbri4.00, lot number: 61p6706, lot quantity: (b)(4) lbs.Inspection instruction met all inspection acceptance criteria.Certified test report dated 05-jun-2020 was reviewed and determined to be conforming.Lot summary report dated 09-jul-2020 met all inspection acceptance criteria.Raw material receiving/putaway checklist met all inspection acceptance criteria.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRANIAL-SCR PLUSDRIVE Ø1.6 SELF-DRILL L4
Type of Device
PLATE, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14840171
MDR Text Key302484894
Report Number8030965-2022-04383
Device Sequence Number1
Product Code JEY
UDI-Device Identifier07611819753457
UDI-Public(01)07611819753457
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K022012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number400.834
Device Lot Number149P092
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/01/2022
Initial Date FDA Received06/28/2022
Supplement Dates Manufacturer Received07/18/2022
08/02/2022
Supplement Dates FDA Received07/22/2022
08/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/05/2021
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-