BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGE; SALINE VASCULAR ACCESS FLUSH Back to Search Results
Model Number 306546
Device Problem Fungus in Device Environment (2316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2022
Event Type  malfunction  
Event Description
It was reported that mold was found around 2 of the bd posiflush¿ normal saline syringes' plunger rods.The following information was provided by the initial reporter: "mold seen around the plunger rod of the flush syringe".
 
Manufacturer Narrative
Investigation summary: it was reported there was mold seen around the plunger rod of the flush syringe.To aid in the investigation, one photo was provided for evaluation by our quality team.The photo shows two syringes in their packaging flow wrap.From the photo it appears that one of the flushes has embedded degraded resin.No other defects or imperfections were observed.A device history record review was completed for provided material number 306546, lot number 1335315.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.Based on the investigation and with the photo sample analysis the symptom reported by the customer is confirmed, but without the physical sample analysis a probable root cause could not be offered.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL SALINE SYRINGE
Type of Device
SALINE VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14844878
MDR Text Key297077292
Report Number1911916-2022-00326
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number306546
Device Catalogue Number306546
Device Lot Number1335315
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/01/2022
Initial Date FDA Received06/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-