• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-S
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2022
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to olympus for evaluation.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus, during reprocessing, the endoscope reprocessor displayed error code e99, indicating that many days have passed since the replacement of the disinfectant.It was noted that the disinfectant will be replaced.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was not returned; therefore, the reported phenomenon or condition of the device could not be confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It was confirmed that the device was shipped according to the specifications.It has been over 1 year since the subject device was manufactured.The specific root cause of the reported problem could not be determined at this time because the device was not returned.The following information is stated in the instructions for use (ifu): ¿chapter 3 inspection before use.Checking the disinfectant solution concentration level.On reprocessing an endoscope, always use an acecide checker or a portable concentration checker to check that the disinfectant solution is at an effective concentration.Please be sure to replace the disinfectant solution before the disinfectant effect disappears.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14850637
MDR Text Key302324462
Report Number8010047-2022-10999
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/01/2022
Initial Date FDA Received06/29/2022
Supplement Dates Manufacturer Received09/05/2022
Supplement Dates FDA Received09/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/18/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-