• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MORTARA INSTRUMENT ELI380 ERGO WAM WLAN USB DICOM SEC AHA B; ELECTROCARDIOGRAPH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MORTARA INSTRUMENT ELI380 ERGO WAM WLAN USB DICOM SEC AHA B; ELECTROCARDIOGRAPH Back to Search Results
Model Number ELI380-DCS11
Device Problem Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The eli 380 device is indicated for use to acquire, analyze, display, and print electrocardiograms.Device is indicated for use to provide interpretation of the data for consideration by a physician.Device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.It is not intended as a sole means of diagnosis.The interpretations of ecg offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.Device is indicated for use on adult and pediatric populations.The device is not intended to be used as a vital signs physiological monitor.Eli 380s with the laird wi-fi module will occasionally disassociate from an access point and will not re-associate with an access point until the wi-fi module is reset.During the periods when not associated with an access point the cardiograph is not able to transmit ecgs to an emr and is not able to download orders from an emr.If the eli 380 device loses connection to the wireless network, the device is unable to transmit or receive ekg¿s.Based on an analysis of the risk file, a failed connection on an eli380 may result in a delay of patient treatment which may cause or contribute to a serious injury or death.Therefore, hillrom is reporting this alleged connection failure as a product malfunction.Software upgrades are being performed on-site to resolve the disconnect issue.Based on this information, no further action is required at this time.
 
Event Description
The customer reported they have wireless network disconnect issues with their device at different locations.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELI380 ERGO WAM WLAN USB DICOM SEC AHA B
Type of Device
ELECTROCARDIOGRAPH
Manufacturer (Section D)
MORTARA INSTRUMENT
7865 n 86th street
milwaukee WI 53224
Manufacturer Contact
keighley crosthwaite
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key14850811
MDR Text Key303086166
Report Number2183461-2022-00029
Device Sequence Number1
Product Code DPS
UDI-Device Identifier00812345026917
UDI-Public812345026917
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberELI380-DCS11
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/16/2022
Initial Date FDA Received06/29/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-